PN B05529AE
2-2
Operation
Run Procedure
WARNING
Normal operation may involve the use of solutions and test samples that are
pathogenic, toxic, or radioactive. Operator error or tube failure may generate
aerosols. Do not run potentially hazardous materials in this centrifuge unless all
appropriate safety precautions are taken. Always use the appropriate rotors and
adapters.
Handle all infectious samples according to good laboratory practices and methods
to prevent the spread of disease. Ask your laboratory safety officer to advise you
about the level of containment required for your application and about the proper
decontamination or sterilization procedures to follow if fluids escape from
containers. Biosafe containment should be used when Risk Group II materials (as
identified in the World Health Organization
Laboratory Biosafety Manual
) are
handled; materials of a higher group require more than one level of protection.
Because spills may generate aerosols, observe proper safety precautions
for aerosol containment.
WARNING
The centrifuge must not be used in the vicinity of flammable liquids or vapors, and
such materials should not be run in the centrifuge. During operation you should
come within the 7.6 cm (3-in.) clearance envelope only to adjust instrument
controls, if necessary. Never bring any flammable substances within the 30-cm
(1-ft) area surrounding the centrifuge. Do not lean on the centrifuge or place
items on the centrifuge while it is operating.
Run Procedure
The following detailed operating procedures are summarized at the end of this section. If you
are an experienced user of this centrifuge, you can turn to the summary for a quick review of
operating steps.
Preparation and Loading
For fast temperature equilibration, cool or warm the rotor to the required temperature before the run.
NOTE
For high-speed runs at temperatures of 20°C or higher, prime the refrigeration system (refrigerated
models) by running the instrument at 10°C for 5 to 10 minutes beforehand to prevent overheating.
NOTE
Before installing the rotor, lubricate it following the instructions in the rotor manual.
1
Check the name rating plate for the correct voltage, then plug the power cord into the wall
receptacle.
Содержание Allegra X-30 Series
Страница 4: ...PN B05529AE iv Revision History ...
Страница 8: ...PN B05529AE viii Safety Notice Chemical and Biological Safety ...
Страница 14: ...xiv Tables ...
Страница 18: ...PN B05529AE xviii Introduction CE Mark ...
Страница 30: ...PN B05529AE 1 12 Description Available Rotors ...
Страница 56: ...PN B05529AE 4 6 Care and Maintenance Supply List ...
Страница 62: ...PN B05529AE Warranty 2 Beckman Coulter Inc Allegra X 30 Series Centrifuge Warranty ...
Страница 63: ......