PN B05529AE
1-6
Description
Controls and Indicators
Figure 1.5
Program Keys
▲
▼
(cursor keys)
The cursor keys are up and down arrow keys (
▲
and
▼
), which can be pressed to
increment values up or down when setting parameters.
ENTER
Parameter (speed, time, temperature, and acceleration or deceleration curve)
changes made while a run is in progress must be verified by pressing the
ENTER
key.
RPM
When the
RPM
key is pressed the last digit in the
SPEED
display (
0
) flashes, indicating
that the speed can be entered in increments of 100 revolutions per minute (rpm).
After the run starts, the actual rpm of the rotor is displayed.
RCF
The
RCF
key can be used to select the speed setting by required relative centrifugal
field (RCF). The corresponding rpm is automatically calculated and displayed during
the run. If the
RCF
key is pressed during the run, the RCF value is shown in the
SPEED
display.
ROTOR
The centrifuge memory contains a list of the rotors that can be used, together with
default parameters for each rotor. When the
ROTOR
key is pressed the number of the
rotor used in the previous run is shown on the
SPEED
display. The rotor list can be
scrolled through, using the cursor keys, until the required rotor number appears.
TIME
The
TIME
key is used to select the run duration. When the
TIME
key is pressed, the
last digit on the
TIME
display flashes, indicating that the time can be entered with the
cursor keys.
•
Timed run — Run time up to 9 hours and 59 minutes can be set. If the minutes
parameter exceeds 59, it is automatically converted into hours.
•
Continuous run — If a run time of less than 1 minute or more than 9 hours and
59 minutes is selected, continuous operation is activated. Time is not counted
down, and the run continues until the
STOP
key is pressed.
TEMP
(refrigerated
models only)
The
TEMP
key is used to select run temperature on refrigerated model centrifuges.
When the
TEMP
key is pressed, the
TEMP°C
display flashes, indicating that the
temperature can be entered with the cursor keys. Temperature can be set between
–20 and +40°C. The operating temperature range is +2°C to +40°C, depending on
the rotor used and the speed selected.
ACCEL
The
ACCEL
key is used to select acceleration rates that protect delicate gradients.
When the
ACCEL
key is pressed, the
ACC/DEC
display flashes, indicating that one of
ten preset rates can be entered with the cursor keys (9 is the fastest rate and 0 is the
slowest rate). Acceleration rates are described in
Table 2.1
in
CHAPTER 2,
Operation
.
DECEL
The
DECEL
key is used to select deceleration rates that maintain optimum separation
while protecting delicate gradients. When the
DECEL
key is pressed, the
ACC/DEC
display flashes, indicating that one of ten preset rates can be entered with the cursor
keys (9 is the fastest rate and 0 is a no-brake coast to stop). Deceleration rate
selections are described in
Table 2.1
in
CHAPTER 2,
Operation
.
Содержание Allegra X-30 Series
Страница 4: ...PN B05529AE iv Revision History ...
Страница 8: ...PN B05529AE viii Safety Notice Chemical and Biological Safety ...
Страница 14: ...xiv Tables ...
Страница 18: ...PN B05529AE xviii Introduction CE Mark ...
Страница 30: ...PN B05529AE 1 12 Description Available Rotors ...
Страница 56: ...PN B05529AE 4 6 Care and Maintenance Supply List ...
Страница 62: ...PN B05529AE Warranty 2 Beckman Coulter Inc Allegra X 30 Series Centrifuge Warranty ...
Страница 63: ......