Installation
Transportation Safety Devices
a.
Make sure that the centrifuge front feet are fully supported on the table.
b.
Locate the centrifuge in an area with sufficient ventilation to allow for heat dissipation.
c.
Check that there are 7.6-cm (3-in.) clearances at the sides of the centrifuge to ensure
sufficient air circulation.
Dimensions are shown in
Figure A.1
. Additional clearance is required on the right side to
allow access to the power switch.
The centrifuge must have adequate air ventilation to ensure compliance to local
requirements for vapors produced during operation.
Ambient temperatures during operation should not be lower than 10°C (50°F) or higher
than 35°C (95°F) for refrigerated models, or lower than 4°C (39.2°F) or higher than 35°C
(95°F) for non-refrigerated models. Relative humidity should not exceed 80%
(noncondensing).
NOTE
During transport between areas with varying temperatures, condensation may occur inside
the centrifuge. Allow sufficient drying time before running the centrifuge.
Transportation Safety Devices
A sheet of foam rubber is installed in the rotor chamber at the factory. A hole in the center of the
foam stabilizes the drive shaft during transport. On receipt of the centrifuge, remove the foam and
store it in case future relocation of the centrifuge is necessary.
Electrical Requirements
See the
Nonrefrigerated Model Specifications
or
Refrigerated Model Specifications
in
CHAPTER 1
for
centrifuge electrical requirements.
Make sure the voltage and frequency imprinted on the name rating plate affixed to the back of the
centrifuge agree with the line voltage and frequency of the outlet used. Refrigeration does not
function properly if the frequency [Hz] does not match the name rating plate. Plug in both ends of
the centrifuge power cord. If there is any question about voltage, have a qualified service person
measure it under load while the drive is operating.
A 1.8-m (6-ft) power cord with grounded plug is supplied with the centrifuge. Make sure that the
matching wall outlet is located near the centrifuge and is easily accessible.
NOTE
The power plug serves as the Disconnecting Device and must remain easily accessible.
Содержание Allegra X-30 Series
Страница 4: ...PN B05529AE iv Revision History ...
Страница 8: ...PN B05529AE viii Safety Notice Chemical and Biological Safety ...
Страница 14: ...xiv Tables ...
Страница 18: ...PN B05529AE xviii Introduction CE Mark ...
Страница 30: ...PN B05529AE 1 12 Description Available Rotors ...
Страница 56: ...PN B05529AE 4 6 Care and Maintenance Supply List ...
Страница 62: ...PN B05529AE Warranty 2 Beckman Coulter Inc Allegra X 30 Series Centrifuge Warranty ...
Страница 63: ......