PN B05529AE
3-4
Troubleshooting
Emergency Access
Emergency Access
If the facility power fails only momentarily, the centrifuge resumes operation when power is
restored and the rotor returns to set speed. However, if the rotor comes to a complete stop you have
to restart the run when the power is restored. In the event of an extended power failure, you may
have to trip the door-locking mechanism manually to remove the rotor and retrieve your sample.
WARNING
Any maintenance procedure requiring removal of a panel exposes the operator to
the possibility of electrical shock and/or mechanical injury. Therefore, turn the
power off and disconnect the instrument from the main power source by removing
its power plug from the receptacle, and refer such maintenance to qualified
service personnel.
1
Turn the power switch to off (
O
) and disconnect the power cord from the main power source
by removing its power plug from the receptacle.
WARNING
Never attempt to override the door interlock system while the rotor is spinning.
2
Make sure that the rotor is not spinning.
CAUTION
Do not proceed if there is any sound or vibration coming from the drive.
3
Use a small flat-blade screwdriver to pry off the two round screw-hole covers on the centrifuge
front panel (see
Figure 3.1
).
Figure 3.1
Screw Locations
1.
Upper screws and covers.
2.
Lower screws, not visible from the front.
Содержание Allegra X-30 Series
Страница 4: ...PN B05529AE iv Revision History ...
Страница 8: ...PN B05529AE viii Safety Notice Chemical and Biological Safety ...
Страница 14: ...xiv Tables ...
Страница 18: ...PN B05529AE xviii Introduction CE Mark ...
Страница 30: ...PN B05529AE 1 12 Description Available Rotors ...
Страница 56: ...PN B05529AE 4 6 Care and Maintenance Supply List ...
Страница 62: ...PN B05529AE Warranty 2 Beckman Coulter Inc Allegra X 30 Series Centrifuge Warranty ...
Страница 63: ......