PN B05529AE
2-3
Operation
Run Procedure
2
2
Press the power switch to on (
I
).
3
Press the
OPEN DOOR
key and lift the door up; it remains in the open position.
4
Use the T-handle wrench to turn the rotor tie-down screw to the left (counterclockwise).
a.
Remove the tie-down screw.
5
Make sure that the tapered sleeve is in place at the base of the centrifuge drive shaft before
installing the rotor (see
Figure 2.1
) and wipe the sleeve to be sure that it is clean and dry.
•
The rotor rests on the sleeve while spinning, and does not operate properly if the sleeve is
missing.
Figure 2.1
Tapered Sleeve Position
NOTE
If the tapered sleeve comes off, it must be replaced by a Beckman Coulter Field Service
representative. Call Beckman Coulter Field Service (1-800-742-2345 in the United States; outside
the U.S. contact your local Beckman Coulter office or visit us at www.beckmancoulter.com).
CAUTION
Do not drop the rotor onto the drive shaft. The shaft can be damaged if the rotor
is forced sideways or dropped onto it. Install the rotor by centering it over the
shaft and carefully lowering it straight down.
6
Install the rotor according to the instructions in the applicable rotor manual.
For a swinging bucket rotor, the steps include the following:
1.
Tapered Sleeve
1
Содержание Allegra X-30 Series
Страница 4: ...PN B05529AE iv Revision History ...
Страница 8: ...PN B05529AE viii Safety Notice Chemical and Biological Safety ...
Страница 14: ...xiv Tables ...
Страница 18: ...PN B05529AE xviii Introduction CE Mark ...
Страница 30: ...PN B05529AE 1 12 Description Available Rotors ...
Страница 56: ...PN B05529AE 4 6 Care and Maintenance Supply List ...
Страница 62: ...PN B05529AE Warranty 2 Beckman Coulter Inc Allegra X 30 Series Centrifuge Warranty ...
Страница 63: ......