PN B05529AE
2-8
Operation
Run Procedure
Setting RCF
1
Press the
RCF
key.
•
The last digit on the
SPEED
display (
0
) flashes, indicating that you can enter the RCF.
2
Press the
▲
or
▼
cursor key until the required RCF is displayed.
•
The corresponding RPM is automatically calculated and the centrifuge runs at the
calculated speed.
Setting Run Time
Run time can be set for either a timed run or continuous operation.
Timed Run
Time can be set for up to 9 hours and 59 minutes (if the minutes parameter entered exceeds 59, it is
automatically converted into hours). During centrifugation, the
TIME
display begins counting down
when the rotor starts to spin and continues the count-down until deceleration begins. The
TIME
display shows the time remaining in the run, in hours and minutes. When the time display reaches
zero, the run ends.
Continuous Run
If a run time of less than 1 minute or more than 9 hours and 59 minutes is selected, continuous
operation is activated. Time is not counted down during continuous operation; instead, the infinity
(
) symbol, indicating continuous operation, lights up and time elapsed since the run start is
displayed. The run continues until the
STOP
key is pressed.
1
Press the
TIME
key.
The last digit on the
TIME
display flashes, indicating that the time can be entered with the
cursor keys.
2
Press the
▲
or
▼
cursor key until the required run duration is displayed.
Setting Run Temperature (refrigerated models only)
Run temperature can be set between –20 and +40°C. The typical operating range is from +2°C to
40°C, depending on the rotor and speed selected.
NOTE
Temperatures may vary slightly between instruments. If sample temperature is crucial, test
temperature settings on your instrument using water samples.
1
Press the
TEMP
key.
Содержание Allegra X-30 Series
Страница 4: ...PN B05529AE iv Revision History ...
Страница 8: ...PN B05529AE viii Safety Notice Chemical and Biological Safety ...
Страница 14: ...xiv Tables ...
Страница 18: ...PN B05529AE xviii Introduction CE Mark ...
Страница 30: ...PN B05529AE 1 12 Description Available Rotors ...
Страница 56: ...PN B05529AE 4 6 Care and Maintenance Supply List ...
Страница 62: ...PN B05529AE Warranty 2 Beckman Coulter Inc Allegra X 30 Series Centrifuge Warranty ...
Страница 63: ......