PN B05529AE
4-2
Care and Maintenance
Cleaning
3
Use a sponge to wipe condensation out of the chamber bowl between runs to prevent chamber
icing (refrigerated models).
a.
If chamber icing occurs, defrost before use.
4
To prevent the rotor from sticking, lubricate the drive shaft with Spinkote at least once a
month, and after each cleaning.
Cleaning
Frequent cleaning is recommended to prolong the life of the centrifuge. Always clean up spills when
they occur to prevent corrosives or contaminants from drying on component surfaces.
NOTE
Before using any cleaning or decontamination methods except those recommended by the
manufacturer, users should check with the manufacturer that the proposed methods will not damage the
equipment.
1
To prevent accumulations of sample, dust, and/or glass particles from broken sample tubes,
keep the interior of the rotor chamber clean and dry by frequent wiping with a cloth or
paper towel.
2
Clean the drive shaft, shaft cavity, threads, and the tie-down screw at least once a week using a
mild detergent such as Beckman Solution 555 and a soft brush.
a.
Dilute the detergent 10 to 1 with water.
b.
Rinse thoroughly and dry completely.
c.
Lubricate the drive shaft with Spinkote after cleaning.
3
Wash the bowl using a mild detergent such as diluted Solution 555.
a.
Rinse thoroughly and dry completely.
b.
If a cleaning solution other than Solution 555 is used, consult
Chemical Resistances
(publication IN-175) or contact the cleaning-solution vendor to verify that the solution
will not damage the centrifuge.
4
Clean the centrifuge case and door by wiping with a cloth dampened with diluted Solution 555.
Do not use acetone or other solvents.
Содержание Allegra X-30 Series
Страница 4: ...PN B05529AE iv Revision History ...
Страница 8: ...PN B05529AE viii Safety Notice Chemical and Biological Safety ...
Страница 14: ...xiv Tables ...
Страница 18: ...PN B05529AE xviii Introduction CE Mark ...
Страница 30: ...PN B05529AE 1 12 Description Available Rotors ...
Страница 56: ...PN B05529AE 4 6 Care and Maintenance Supply List ...
Страница 62: ...PN B05529AE Warranty 2 Beckman Coulter Inc Allegra X 30 Series Centrifuge Warranty ...
Страница 63: ......