PN B05529AE
2-14
Operation
Summary of Run Procedures
4
Close the centrifuge door and push firmly down on it until you hear the latch engage.
5
Enter run parameters:
a.
Select a rotor number —
ROTOR
,
▲
or
▼
,
ENTER
b.
Set run speed —
RPM
,
▲
or
▼
;
or
RCF
,
▲
or
▼
c.
Set run duration —
TIME
,
▲
or
▼
d.
Set run temperature —
TEMP
,
▲
or
▼
e.
Select acceleration rate (0 through 9)—
ACCEL
,
▲
or
▼
f.
Select deceleration rate (0 through 9)—
DECEL
,
▲
or
▼
6
Check that all parameters are correct and the door is shut and latched, then press
ENTER
,
then
START
.
WARNING
Never attempt to override the door interlock system while the rotor is spinning.
CAUTION
Do not lift or move the centrifuge while the rotor is spinning.
7
Wait for the set time to count down to zero, or end the run by pressing or holding the
STOP
key.
8
After the rotor stops spinning
and the
OPEN DOOR
light comes on, press the
OPEN DOOR
key to
release the door latch; open the door.
9
Unload the rotor according to instructions in the applicable rotor manual.
CAUTION
If disassembly reveals evidence of leakage, you should assume that some fluid
escaped the rotor. Apply appropriate decontamination procedures applied to the
centrifuge and accessories.
Содержание Allegra X-30 Series
Страница 4: ...PN B05529AE iv Revision History ...
Страница 8: ...PN B05529AE viii Safety Notice Chemical and Biological Safety ...
Страница 14: ...xiv Tables ...
Страница 18: ...PN B05529AE xviii Introduction CE Mark ...
Страница 30: ...PN B05529AE 1 12 Description Available Rotors ...
Страница 56: ...PN B05529AE 4 6 Care and Maintenance Supply List ...
Страница 62: ...PN B05529AE Warranty 2 Beckman Coulter Inc Allegra X 30 Series Centrifuge Warranty ...
Страница 63: ......