3 . R e p a i r I n s t r u c t i o n s
1 / 2 0 1 0
3 - 8
BA-TE-DE08C M.KAY
SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
3.2.3
Example LLC/LLS Test Codes
for Blood Leak Detector
L L C T e s t C o d e :
1 1 ( T T ) :
Blood leak detector test
1 0 ( S S ) :
Send signal to blood leak task
2 (x):
Third repetition (x = 0, 1, 2)
L L S T e s t C o d e :
1 1 ( T T ) :
Blood leak detector test
0 0 4 ( S S ) :
Measure level for red and green
LLC Test
Code
LLC Test Steps
LLS Test Steps
LLS Test
Code
1.
1102x
Prerequisites for Test:
Upline tank must be filled
Timeout 2 min
2.
Test code =
11
to LLS
Test code =
11
back to LLC
3.
11
10
x
EV_TESTE_BLUTLECK_EIN to BlutleckTaskID
Test initialisation (VGD ON)
Single measurement of green,
Single measurement of red,
then comparison of: red
≥
green + 0.2 V
no timeout
11000 –
11003
11
004
4.
Acknowledge with switch-off of VGD
11005 –
5.
1120x
Wait for D24OFF (timeout 30 s)
11011
6.
VGD ON
11012
7.
1130x
Wait for D24ON (timeout 30 s)
8.
1199x
Completed
3.2.4
Overview Test Codes
0 0
No tests
1 1
Blood leak detector
1 2
DFS pressure test
1 5
Integrity test HDF online (only for HDF online)
1 6
Alarm system
1 7
Monitoring of analog voltage
±
12 V
2 0
UF pump
2 1
End conductivity
2 2
Temperature TSD
2 5
SMPS-MC EEPROM
2 6
SMPS-MC service mode
2 7
SMPS-MC battery
2 8
SMPS-MC buzzer
3 0
SAD: Level test
3 1
SAD: Counter test
3 2
Pressure sensors blood side: Equality test for blood side pressure sensors
3 3
Pressure sensors blood side: Blood side pressure test
3 7
Disinfection Valve VD
4 0
Pressure test substitution line S-Online HDF online
1 1
1 0
2
1 1