3 . R e p a i r I n s t r u c t i o n s
1 / 2 0 1 0
3 - 87
BA-TE-DE08C M.KAY
SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
3.6.10.3 Read All Trends
A separate software program is required to analyse the trends, i.e. the
Trend Viewer.
The utility reads all possible backup trend files on the hard disk drive
(CFC) (max. 20 trends) and writes them to the USB stick. Files from different
machines are placed in different subdirectories on the USB stick. The
subdirectory name USER/12_Read_All_Trends/ <SERNUM>/ is created from the
serial number of the machine. Files from the same machine are placed
in different subdirectories where the subdirectory names are created from the
trend creation dates <DATE>.
<SERNUM>
The subdirectory name is created from the serial number <SERNUM> of
the . If the serial number is not set on the <SERNUM> is
equal to ‘-----‘.
<DATE>
The subdirectory name is created from a date. Format: YYYY-MM-
DD.hh.mm.
In case of TRENDS: the date is the date of the trend start.
12_Read_All_Trends
BlueCat
Removable Disk (X: - FSU USB Stick)
DISKTYPE
USER
05_Write_Calibration_Data
04_Write_Languages
06_Write_System_Configuration
09_Card_01_Read_Card_Image
09_Card_02_Write_Card_Image
09_Card_03_Read_Patient_Diskette
09_Card_04_Write_Patient_Diskette_to_Card
09_Card_05_Read_Patient_Card
10_Read_System_Configuration
09_Card_06_Write_Patient_Card
11_Read_Calibration_Data
See the results in the directories with the following path for the
Read All Trends file tbd1.zip.
Read All Trends
USB:\USER\12_Read_All_Trends\{Serial Number}\{YYYY-MM-
DD.hh.mm}\tbd1.zip
Example:
USB:\USER\12_Read_All_Trends\{Serial Number}\2007-09-
12.14.25}\tbd1.zip
Y Y Y Y :
Year 2007
M M :
09 month September
D D :
12 Wednesday
h h :
14 hours
m m :
25 minutes