4 . T S M S e r v i c e P r o g r a m
1 / 2 0 1 0
4 - 170
BA-TE-DE08W M.KAY
SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
4.8.3.26
Arterial Tubing Clamp SAKA/Venous Tubing
Clamp Currentless Closed SAKV-SG
SAKA
You can check the soft stoppers
A
on the V-stopper/stopper
B
of the
arterial tubing clamp
SAKA
.
•
Soft stoppers art. no. 34570675
1.
Open front door.
2.
Check the soft stoppers
A
on the V-stopper/stopper
B
.
Replace if soft stoppers show any damages (tears/ruptures/
brittleness).
SAKV-SG
You can check the o-rings
C
on the V-stopper/stopper
D
of the
venous tubing clamp currentless closed
SAKV-SG
.
•
O-ring 7 x 1.5 art. no. 34570624
3.
Check the o-rings
C
on the V-stopper/stopper
D
Replace if o-rings show any damages (tears/ruptures/
brittleness).
4.
Close front door.
SAKA
SAKV-SG
B
D
A
C