4 . T S M S e r v i c e P r o g r a m
1 / 2 0 1 0
4 - 133
BA-TE-DE08W M.KAY
SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
1.
Activate the LL Manual Calibration menu with
the LL Manual Calibration key in the Manual
Test and Calibration menu.
2.
Activate the Dialysis Temperature menu by
pressing on the line Calibration 2.5 Dialysis
Temperature. The following menu is displayed.
•
Measurement instrument (e.g. HDM 99 or equivalent)
•
Adapter set (coupling, tubing, dialyser couplings)
3.
Remove
TSDE
from rinsing bridge and connect at the outlet of
FPE
(tubing going to
FBK1
).
Note:
Pay attention to correct flow direction of temperature sensor
TSDE
.
4.
Connect measurement instrument between
TSDE
and
FBK1
.
5.
Connect dialyser couplings to flow meter.
6.
Increase slowly the speed of
FPE
, until a flow of 750 ml/min
±
50 ml/min
is reached.
7.
Press
key for a combined calibration of
TSBIC
,
TSD
/
TSD-S
and
TSDE
(The circulation valve
VZ
stays closed).