4 . T S M S e r v i c e P r o g r a m
1 / 2 0 1 0
4 - 123
BA-TE-DE08W M.KAY
SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
4.8.3.5.1
Calibration Degassing Pressure
Sensor PE for Machines at High
Altitudes of >1,000 m
For calibration point -400 mmHg/-530 mbar
If the Dialog machine is used at high altitudes (>1,000 m above sea level), an
offset of +75 mmHg/+100 mbar per 1000 m must be used for the calibration
of the degassing pressure sensor
PE
.
Measurement unit mmHg
1,000 m altitude: –325 mmHg (–400 mmHg + 75 mmHg offset = –325 mmHg)
2,000 m altitude: –250 mmHg (–400 mmHg + 150 mmHg offset = –250 mmHg)
3,000 m altitude: –175 mmHg (–400 mmHg + 225 mmHg offset = –175 mmHg)
Measurement unit mbar
1,000 m altitude: –430 mbar (–530 mbar + 100 mbar offset = –430 mbar)
2,000 m altitude: –330 mbar (–530 mbar + 200 mbar offset = –330 mbar)
3,000 m altitude: –230 mbar (–530 mbar + 300 mbar offset = –230 mbar)
Example for Calibration at 1,000 m
Altitude
1.
Clamp tubing at filter degassing pump
FEP
.
2.
Clamp tubing between degassing bypass valve
VEB
and degassing
pressure sensor
PE
and pull off tubing from the degassing pressure sensor
PE
.
3.
Connect a manometer and a syringe to the
PE
sensor, using a Y-piece and
extra tubing.
PE Calibration Point -400 mmHg
–530 mbar
4.
Press the Degassing Pressure PE
key.
5.
Use syringe and create a negative pressure, e.g. at 1,000 m altitude:
•
–325 mmHg
±
10 mmHg (-400 mmHg plus 75 mmHg/1,000 m)
•
-430 mbar
±
13 mbar (-530 mbar plus 100 mbar/1,000 m)
6.
Press
key.
7.
Enter the stable reference value:
•
–325 mmHg
±
10 mmHg (-400 mmHg plus 75 mmHg/1,000 m)
•
–430 mbar
±
13 mbar (-530 mbar plus 100 mbar/1,000 m)
Wait until digits are stable.
8.
Confirm with
key.
PE Calibration Point 0 mmHg
0 mbar
11.
Remove syringe (0 mmHg/0 mbar).
12.
Press
key.
13.
Enter 0 mmHg/0 mbar.
Wait until digits are stable.
14.
Confirm with
key.
15.
Confirm and store complete calibration procedure with
key.
Note:
The calibration data is saved only after the bit sequence 0-1-0 (for
detailed information see 4.8.3.1 General Calibration Information).
16.
Remove calibration devices.
17.
Reconnect tubings.