4 . T S M S e r v i c e P r o g r a m
1 / 2 0 1 0
4 - 52
BA-TE-DE08C M.KAY
SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
4.7.1.17
bioLogic RR
®
GL Parameters
1.
Activate the System Configuration menu with
the System Configuration key in menu
Treatment Support. The following menu is
opened and displayed (scroll with scroll bar).
2.
Activate the bioLogic RR
®
Comfort menu with
the bioLogic RR
®
Comfort icon. The following
menu is opened and displayed.
bioLogic RR
®
GL
(GL: Guide-Line)
Note:
bioLogic RR GL = bioLogic RR Comfort
The parameters for the bioLogic RR
®
Comfort can
be set in the System Configuration, bioLogic RR
®
Parameters menu, if the option was already
selected and activated in the Production Report
menu.
The following parameters can be set:
•
Guide Line Mode
•
Suggested SYS (Systolic) Lower Limit
•
bioLogic RR Algorithm ID
Guide Line Mode
Press YES key to enable the Guide Line (GL) mode of the bioLogic RR
®
Comfort
option.
If NO is selected the bioLogic RR
®
Comfort option is used without the guide
line (GL) mode.
Suggested SYS (Systolic) Lower Limit
If Guide Line Mode: YES
The suggested systolic lower limit can be selected. The suggested systolic lower
limit can be used in therapy to set the SYS lower limit.
If Guide Line Mode: NO
The suggested systolic lower limit can not be selected (see bottom figure).
bioLogic RR Algorithm ID
ID Number 200
The ID number 200 indicates that the Guide Line Mode is enabled and can be
used in therapy.
ID Number 100
The ID number 100 indicates that the Guide Line Mode is not enabled and can
not be used in therapy.