4 . T S M S e r v i c e P r o g r a m
1 / 2 0 1 0
4 - 27
BA-TE-DE08C M.KAY
SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
Limit Maximum PBE
The maximum limit for the monitoring of PBE in
therapy can be set.
Limit Delta PBE
The permissible changes for PBE during therapy
can be set.
Limit Maximum TMP
The maximum upper TMP limit for the monitoring
in therapy can be set. This value is always active if
no relative monitoring is active (see description
for Limits Low/High).
Limits Low/High TMP
The relative lower and upper limit can be set in %
for TMP monitoring in therapy.
Limit OFF:
Monitoring is not performed
(activated)
Limit OFF:
Monitoring is performed
(deactivated)
Minimum TMP Alarm Window
The limit window is at least as large as predefined
independent of the relative low and high limit.
The default is 20 mmHg and can be changed in
the limits of 20 to 100 mmHg.
3.
Use the scroll bar until the following menu
appears.
Extended TMP Limit Range
Key
is Displayed in Treatment
The user can select the extended TMP range.
NO
The Extended TMP Limit Range key can not be
selected/enabled (shaded) if NO is selected.
YES
The Extended TMP Limit Range key can be
selected/enabled (highlighted) if YES is selected.
Select Extended TMP Limit Range
If the key was selected, the extended TMP limit
range can be preselected during the start of the
therapy via the Extended TMP Limit Range.
If the key is not pressed
The TMP low limit range is set min. –10 mmHg.
If the key is pressed
The TMP limit range is extended to –100 mmHg
(absolute TMP value).