4 . T S M S e r v i c e P r o g r a m
1 / 2 0 1 0
4 - 173
BA-TE-DE08W M.KAY
SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
4.9 Production
Report
Dialog
B.Braun Avitum AG
Hersteller /
SN
VA
REF
V
Hz
IP 21
0123
50/60
230~
710xxx
2500
10xxxx
34209 Melsungen, Germany
Not to be used in presence of
flammable anaesthetic gases
+
Fig.: Type Plate on Rear Door
The delivery data of the Dialog are displayed in the menu, i.e. must be entered
during the commissioning of the machine, e.g.:
•
Type = REF (see type plate)
•
Ser-No. = SN (see type plate)
Options can be activated in the following menus:
Icons
For example: bioLogic RR, card reader or K*t/V-UV (Adimea)
Low Level Options (see 4.9.2)
For example: Double pump or BIC cartridge holder
Activate the Production Report menu with the
Production Report key in TSM Main Menu.
Options assembled in the machine must be
activated. Further options present in the Dialog
must also be activated in this menu or in the Low
Level Options menu (see 4.9.2).
The following options can be activated:
•
ABPM
•
bioLogic RR (icon available if installed)
•
Low Level Options
•
Card reader
•
DCI (Data Communication Interface
•
K*t/V
•
K*t/V-UV (Adimea)
•
HCT (Crit-Line interface)
ABPM Automatic Blood
Pressure Measurement
If the ABPM option is assembled in the machine or the option is retrofitted it
must be selected and saved to activate the option.
bioLogic RR Automatic Blood
Pressure Stabilisation
The bioLogic RR icon is displayed after the installation software was installed
for the bioLogic RR option. The bioLogic RR must be selected and saved to
activate the option.
Card Reader
If the card reader option is present in the machine, it must be selected and
saved to activate the option.
DCI Data Communication
Interface
The DCI option communicates with the Dialog machine via an RS 232
interface. The communication is performed via a DIANET protocol.
K*t/V
The KT/V option must be selected and saved to activate the option.
Adimea (K*t/V-UV)
(
A
ccurate
di
alysis
mea
surement)
If an Adimea option is present in the machine, it must be selected and saved to
activate the option.
HCT Crit-Line
If a Dialog Serial Interface DSI is present in the machine, it must be selected
and saved to activate the option.
Note
•
The activation of the HCT Crit-Line option enables a window in therapy.
This window allows the data exchange and display of measurement values
to a connected HemaMetrics
™
Crit-Line device.