4 . T S M S e r v i c e P r o g r a m
1 / 2 0 1 0
4 - 174
BA-TE-DE08W M.KAY
SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
4.9.1
Dialog Version Number
The Dialog Version Number menu is selected with
the Dialog Version Number window in Production
Report. The current software/firmware versions and
hardware versions are displayed. This menu can also
be selected in the TSM Main Menu.
The various software versions for top level,
controller, supervisor, etc. are displayed and can be
checked. These versions depend on the installed
Dialog software version. The following software and
hardware statuses can be checked:
Component SW Versions
The versions for the installed software/firmware are
displayed.
•
Top Level Controller Software TLC SW
•
System Configuration
(displays the system configuration version
number for the haemodialysis machine)
•
Low Level Controller Software LLC SW
•
Low Level Controller Firmware LLC FW
•
Low Level Supervisor Software LLS SW
•
Low Level Supervisor Firmware LLS FW
•
Automatic Blood Pressure Measurement ABPM
(firmware for the ABPM module is displayed)
Language Versions
The installed languages are displayed.
Treatment
•
044 - English
•
049 - German
•
or further therapy languages
TSM
•
044 – English
•
049 – German
Low Level Hardware Versions
The boards are displayed with the respective
hardware versions, e.g. a HDF online machine. The
displayed HW versions depend on the assembled
boards in the machine.
•
Basic Board
BB
•
Analog Board
AB
•
Power Board Motors
BPM
•
Power Board Valves
BPV
•
HDF Online Board
HOB
•
Digital Board
DB