4 . T S M S e r v i c e P r o g r a m
1 / 2 0 1 0
4 - 58
BA-TE-DE08C M.KAY
SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
Elapsed Months
A warning message is enabled if YES is selected and the preselected
period of months is expired.
Note:
The manufacturer recommends an annual preventive
maintenance and the replacement of the wear and tear
parts from the preventive maintenance kit.
Elapsed WTC
(Working Time Counter)
A warning message is enabled if YES is selected and the set working
time is elapsed.
Preventive Maintenance Accomplished
Press the Preventive Maintenance Accomplished key and then the
Save icon after the preventive maintenance was performed.
Consequently a new preventive service period is started again after
the machine is switched off and on.