3 . R e p a i r I n s t r u c t i o n s
1 / 2 0 1 0
3 - 37
BA-TE-DE08C M.KAY
SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
(Timeout 10 s since start of subcode 400)
16.
3360x
(Error) the test prerequisites from HDF online task are
not given during one of the test steps (e.g.
autopriming, HDF filter rinsing required after LF error).
17.
3370x
(Error) during start up of BPA (subcode 200), PV >
+400 mmHg
18.
3399x
Completed
33041
3.3.18
Disinfection Valve VD
Objective
Function test of the detection of the actual value for LLS.
The disinfection valve VD is opened shortly (t < 3 s) by LLC during the test. The opening
must be detected by LLS.
The test is performed in Preparation if the dialyser couplings are not connected to the
rinsing bridge anymore and the blood side pressure retention test was passed. If the test
is not passed therapy is not possible.
Test code
37
Repeats until error message
3
ID error message
1165: Self test VD failed
Actions after error
None (repeat test until passed).
LLC Test
Code
LLC Test Steps
LLS Test Steps
LLS Test
Code
1.
Prerequisites for Test: None
2.
Test code = 37 to LLS
Test code = 37 back to LLC
3.
3710x
Opened VD
Test initialisation
37000
4.
Start condition for the test:
1.
BS pressure retention test is passed (test code 33)
37001
5.
Detection valve opened
37002
6.
3720x
Close VD after 750 ms
7.
Detection valve closed. The detected opening is
acknowledged by the switch-off of VGB.
37003 –
37009
8.
3730x
Wait for B24OFF (timeout 7.5 s)
9.
VGB ON
37010
10.
3730x
Wait for B24ON (timeout 5 s)
11.
3780x
(Error) at start of test one or both couplings connected
to rinsing bridge.
12.
3799x
Completed