3 . R e p a i r I n s t r u c t i o n s
1 / 2 0 1 0
3 - 81
BA-TE-DE08C M.KAY
SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
3.6.9.8 Remove All Patient Data
Remove all patient data from the FSU USB stick after service with the
utility
07. Remove All Patient Data
. This will delete all patient data stored
on the FSU USB stick.
The service technician can store patient data on the USB stick by reading them
from patient cards and patient diskettes. Additionally patient data can be
copied manually to the 09_Write_Patient_Diskette_to_Card, DEFAULT
subdirectory and the 09_Write_Patient_Card, DEFAULT sub-directories. If the
patient data are not required anymore the service technician can remove them
from the USB stick in one step by using the 07 Remove All Patient Data utility.
The utility will remove all the patient data from the FSU USB stick, except the
patient card images.
The utility requires two confirmations.
First confirmation:
Are you sure?
AQ
: exit from the utility
EQ
: renewed query for confirmation
Second confirmation:
Are you REALLY sure?
BP
:
exit from the utility
BP
:
all patient data are deleted from the FSU USB stick.
F i e l d S e r v i c e U t i l i t i e s V 2 . 0 0 - S e t 0 0 0
CRC = XXXX - OK
01. Read Card Image
(Card
→
USB Stick)
02. Write Card Image
(USB Stick
→
Card)
03. Read Patient Diskette
(Diskette
→
USB Stick)
04. Write Patient Diskette to Card
(USB Stick
→
Card)
05. Read Patient Card
(Card
→
USB Stick)
06. Write Patient Card
(USB Stick
→
Card)
07. Remove All Patient Data
BP-
: move cursor up
BP+
: move cursor down
AQ
: back to previous menu
EQ
:
execute
1.
Press
EQ
to execute the utility.
2.
Press
AQ
to skip to the previous menu.