4 . T S M S e r v i c e P r o g r a m
1 / 2 0 1 0
4 - 22
BA-TE-DE08C M.KAY
SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
Default Conductivity Mode
The bicarbonate dialysis is the standard factory setting. You can preselect
either acetate or bicarbonate for the standard dialysis operation mode.
1.
Click either the Bic or Acet key for the standard setting.
Stand-by Function is Available in
Preparation
The stand-by function is available during preparation.
1.
Click the NO key to disable the stand-by function in preparation.
Stand-by after Self Test/Rinsing
The function is activated after the self test/rinsing.
Maximum Stand-by Time
You can set the maximum duration of the stand-by state.
Selected Acetate
There are 10 acetate concentrate names and the
conversion factors for the conversion of
conductivity values for concentrate of sodium
ions.
If necessary the name and the conversion factor
[mS/cm per mmol/l] can be set, for the selected
acetate type.
1.
Click the ACET icon to activate the Selected
Acetate menu.
The following menu is opened. You can activate a
concentrate for therapy by pressing a free section
of the window.
2.
Click the Concentrate window if necessary to
edit a concentrate name. This name is
displayed in therapy mode.
3.
Click the Conversion window if necessary to
change the conversion factor [mS/cm per
mmol/l].
Conversion Factor:
4.
Click the Default Value window if necessary to
change the value [mS/cm].
Conversion Factor
NaCl Concentration mmol/l (Measured Value)
Conductivity mS/cm (Standard Value)
=