4.8.2
Low Level Manual Test
You can test the function of the low level controller LLC sensors, pumps and
valves. You can select a test menu and check a single (or where applicable
combined) sensor, pump and valve. The components are displayed in graphics
(position and corresponding value is displayed). Additionally you can directly
skip to the Low Level Calibration menu with a Calibration key.
1.
Activate the Manual Test and Calibration
menu with the Manual Test and Calibration
key in the TSM Main Menu. The following
menu is displayed.
2.
Activate the LL Manual Test menu with the LL
Manual Test key in the Manual Test and
Calibration menu. The following menu is
displayed (if necessary use scroll bar).
The following tests can be selected:
•
1.1 Staff Call, Alarms, Power off
•
1.2 Blood Pump Arterial
•
1.3 Blood Pump Venous
•
1.4 Heparin Pump
•
1.5 Tubing Clamps
•
1.6 Pressure Sensors (Blood Side)
•
1.7 Air Sensor, Red Sensor
•
1.8 Coupling Detectors
•
1.9 Water Intake, Upline Tank and Flow Pump FPE
•
1.10 Degassing and Heating
•
1.11 Dialysis Temperature
•
1.12 Concentrate Pumps and Conductivity
•
1.13 Balance Chamber Valves
•
1.14 Bypass and Disinfection Valves
•
1.15 Flow Pump FPA, Balance Chamber and Air
Separator
•
1.16 Pressure Water Part
•
1.17 UF Pump
•
1.18 Blood Leak
•
1.19 Leak Test
•
1.20 Water Part (Overview)
•
1.21 Emptying
•
1.22 HDF
•
1.23 BIC Cartridge Valve and Coupling Detectors
•
1.24 HDF Online
•
1.25 HDF Online Valves and Coupling Detectors
•
1.26 Option Battery
•
1.27 DFS Pressure Test of Therapy
•
1.28 Front Panel Board FPB (LLS)
•
1.29 Patient Leakage Current
•
1.30 SMPS-MC Self Tests
•
1.31 Level Regulation
Function of the Software Keys and Icons
The function of the software keys and icons are as
follows:
Press the HELP key to see the help information of
the menu.
The menu can be closed with the Exit Door key.
If required, the previous menu or the next menu
can be selected with the Previous and Next Arrow
keys.
If required, a corresponding calibration menu can
be selected in certain LL Manual Test menus. The
calibration procedures are described in paragraph
4.8.3 Low Level Manual Calibration.