4 . T S M S e r v i c e P r o g r a m
1 / 2 0 1 0
4 - 34
BA-TE-DE08C M.KAY
SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
Priming without Recirculation
If YES is activated
The blood pump stops 120 ml before the required filling volume. The 120 ml is
used for the blood pump self test. A warning message is not activated. Thus a
connection of the disposable for recirculation is not necessary after filling. An
automatic start of the rinsing is prohibited.
Filling BP Rate
You can set the BP rate to fill the blood line system (BS) for all rinsing modes.
Filling BP Volume
You can set the filling/rinsing volume for the A/V system.
With Rinsing Out of Machine
The filling of the dialyser is started after the filling BP volume is reached (e.g.
500 ml; min. 500 ml/standard 700 ml).
With Bag Rinsing
The blood pump BPA is stopped after the filling BP volume is reached (e.g. 500
ml). This allows the changing from two bags to one bag.
Automatic Rinse Program after Self Tests
If YES is Activated
The rinsing is automatically performed directly after the BS self tests.
If NO is Activated
The rinsing must be started manually after the BS self tests.
User Message for BS Filled?
If YES is Activated
A user message is displayed after the rinsing of the machine:
The BS system is filled until the
key is pressed and the filling BP volume
is completed. Thus the level can be set. Then the dialyser can be turned to fill
the DF side.
If NO is Activated
The filling of the dialyser is automatically started after the Filling BP Volume is
finished.
Rinsing Blood Flow
You can set the blood flow for rinsing. After the self tests BPA delivers a
longitudinal flow during rinsing. This depends on the rinsing mode.
Longitudinal rinsing (along the dialyser):
•
Blood side inlet/outlet
•
Water side inlet/outlet
Cross rinsing (through the dialyser membrane):
•
Blood side to water side
Note
Interdependencies when changing parameters:
Change
Automatic Adaptation from
Rinsing Time
Rate
Rate Rinsing
Time
Rinsing Volume
Rate
Rinsing Dialysate Flow
You can set the dialysate flow for the rinsing of the dialyser in preparation.
Rinsing Time by UFP
You can set the rinsing time (for bag rinsing without substitution line (s-
online)). Thus rinsing via the dialyser membrane is possible. After the
termination of the rinsing time, rinsing is performed with min. UF.
Rinsing Rate by UFP
You can set the rinsing rate for the UFP to reach the Rinsing Volume by UFP.
The Rinsing Time by UFP is automatically recalculated.
Rinsing Volume by UFP
You can set the rinsing volume (via the dialyser membrane) for the Rinsing
Time by UFP. The Rinsing Rate by UFP is automatically recalculated.
Blood Flow for Connecting Patient
You can set the blood flow for connecting the patient (in therapy). The blood
pump BPA is started with this rate when the therapy is started.