7 . S p a r e P a r t s L i s t
1 / 2 0 1 0
7 - 48
Pos. Art. No.
Qty.
Description
Comments
BA-TE-DE08C M.KAY
SW9xx_sm_Chapter 7_1-2010.doc/pdf <100329> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
7.11
Test and Calibration Equipment and Tools
Inspection for Calibration Equipment
# Calibration equipment is subject to inspection and
must be checked and calibrated in regular intervals.
#
7701357
1
Manometer 0-4 bar
#
34511571
1
Manometer 2.5 bar
-
770085A
1
Flow meter 1.1 l/min
-
7700911
1
Reference solution (red)
for calibration blood leak detector
#
7702493
1
Template 1.4/1.5
for tubing clamp
#
-
*
Dialysis measurement
instrument HDM 99XP
For example manufacturer:
IBP Instruments GmbH
Internet: www:ibpmt.com
for temperature, conductivity,
flow and pressure measurement
for details contact manufacturer directly
-
-
-
Crossover cable RS 232
for HDM 99 RS 232 interface
for details contact IBP directly
#
-
*
SecuLife ST
For example manufacturer:
GMC Instruments Group
Internet: www:gossenmetrawatt.com
for electrical safety check according to
IEC 62353/EN 60601-1/IEC 601-1
for details contact manufacturer
directly
-
-
*
Balance (e.g. Pocket Balance)
For example manufacturer:
Kern & Sohn GmbH
Internet: www:kern-sohn.com
for calibration of UFP, BICP and KP
for details contact manufacturer
directly
* not in the product range of B. Braun Avitum AG