2.15 Precautions for Preventing Damage
2-39
2
Before Use
<Recommended procedures>
Measurement Preparation 1:
1. Connect the BERTWave and all equipment.
2. Set the DC power supply output to ON.
3. Set the BERTWave output to ON, and start measurement.
Measurement Preparation 2:
1. Set the BERTWave output to OFF.
2. Set the DC power supply output to OFF.
3. Disconnect the BERTWave and all equipment, or change the DUT
connections.
Since even unforeseen fluctuations in DC voltage and load (open or short
circuits at the BERTWave output side and changes caused by using a
high-frequency probe, etc.) can damage the DUT and the BERTWave, we
recommend connecting a 50
resistance in series with the DC terminal
of the Bias-T to prevent risk of damage.