Chapter 7 Measuring Waveform
7-34
Averaging 1
Averaging 100
Figure 7.7-5 Example of Averaging Process
The averaging is calculated using the following formula.
Number of waveforms
≤
Averaging setting count:
n
n
S
n
Ave
n
n
Ave
1
1
Averaging setting count
≤
Number of waveforms:
M
n
S
n
Ave
M
n
Ave
1
1
Ave(n): Averaging value, S(n): Measurement value, M: Averaging setting
count, n: Number of waveforms
The size of the noise is inversely proportional to the square root of the
averaging setting count.
For example, if M is set to 100, the noise amplitude is compressed to
about 1/10 compared to when M is set to 1.