10.6 Touch Panel Position Calibration
10-11
10
Mai
n
tenance
10.6 Touch Panel Position Calibration
Touch panel position may be misaligned depending on the changes of
ambient environment (temperature/humidity). If the touch panel position
is misaligned, calibrate the position using the following procedures.
When calibrating the detecting position, use the sharp-nose stick but do
not scratch the panel surface.
1. The Windows Control Panel is displayed.
2. Touch [Touch Panel] twice.
The property window of the touch panel is displayed.
3. Touch
[Calibration].
The screen to input the position is displayed.