3.3 Simultaneously Measuring Bit Error Rate and Waveform
3-13
3
Connecting with DUT
CAUTION
Check the following before connecting the optical fiber:
The output power for the optical module inserted
with the Option 051 is within the input level range
for the DUT.
The DUT optical output level does not exceed the
rated optical input level for O/E Data In.
There is a risk of damaging the DUT or optical module
if the optical power exceeding the rated optical input
level is applied.
ED2
Optical Fiber
Coaxial
Cable
Tx Data In
O/E Data In
Optical Attenuator
Optical
Power Meter
PPG1
PPG2
DUT
Trigger Clk Input
Sync Out
Optical coupler
or optical switch
Figure 3.3-1 When Measuring Bit Error Rate and Waveform
Simultaneously (Option 012, 023, and 051)