7.6 Setting Clock Recovery and Rate
7-27
7
Measurin
g W
aveform
CAUTION
●
Set the amplitude of the signal input to the CRU In
connector to not exceed the following values. This is
equivalent to + 10 dBm for a sine-wave signal. Inputting
a signal with a larger voltage risks damaging the
internal circuits.
Option 053: 2 Vp-p
Option 054: –1 dBm (mean)
Option 055: 1.0 Vp-p
Procedure
1. Connect the Trigger CLK In and CRU Out connectors on the front
panel using a coaxial cable.
2. With the Option 053, input the signal to the front panel CRU In
connector.
With the Option 054, input the signal to the front panel O/E Data In
connector.
With the Option 055, input the signal to the front panel ED1-Data
CRU In connector.
3. Touch [EYE/Pulse Scope].
4. Touch
[Setup].
5. Touch the Clock Recovery button to select CRU Band from the
following items.
[Off]:
Turn off the clock output to CRU Out connector.
[<2.7]:
Sets the CRU from 0.1 to 2.7 GHz.
[>8.5]
Sets the CRU from 8.5 to 12.5 GHz.
When setting to [<2.7] or [>8.5], “*CRU Valid” is displayed.
6. When Clock Recovery is [>8.5], select the CRU loop BW from the
following:
[1 MHz], [2 MHz], [4 MHz], [8 MHz]
7. Touch
[Time].
8. Set the divide rate to 1.
9. Touch [Acquire Clock Rate] at Acquire Clock Rate.
Note:
When not using the clock recovery unit, set Clock Recovery to [OFF].
The clock recovery may not be generated if the input signal
frequency to the clock recovery is out of CRU Band. In that case,
the [PLL Unlocked] is displayed.
Содержание BERTWave MP2100B
Страница 24: ...xxiv...
Страница 96: ...Chapter 2 Before Use 2 42...
Страница 112: ...Chapter 3 Connecting with DUT 3 16...
Страница 250: ...Chapter 7 Measuring Waveform 7 64 Figure 7 12 1 Marker Display...
Страница 262: ...Chapter 7 Measuring Waveform 7 76...
Страница 276: ...Chapter 8 Operating Optical Interface 8 14...
Страница 304: ...Chapter 9 Performance Test 9 28...
Страница 313: ...10 4 Displaying Software Version 10 9 10 Maintenance 4 Touch Exit and then the software version display screen is closed...
Страница 320: ...Chapter 10 Maintenance 10 16...
Страница 374: ...Appendix E Performance Test Record Form E 10...
Страница 382: ...Index Index 6...