7.7 Setting Data Collection Method
7-33
7
Measurin
g W
aveform
Figure 7.7-4 Accumulation Type Settings (Infinite)
Displaying averaging waveform
When setting the display method to [Averaging] at the pulse mode, the
averaging process is performed. Averaging process is used to suppress
waveform noise.
1. Touch [Setup] to open the Setup dialog box.
2. Touch the Accumulation Type button to select [Averaging].
3. Touch the Number of samples button to select the number of data
displayed on the screen from the following:
[512], [1024], [2048], [4096], [8192],[16384]
4. Touch the Averaging text box to set the number of waveforms to
perform averaging.
In averaging, the mean value of the data count input at Averaging is
calculated and the results are displayed on the screen. However, when 1 is
input, no averaging is performed.
The number of averaging waveforms is displayed on the screen while
executing averaging.
The waveform is captured ten times when the following condition is set,
and ten averaging waveforms are displayed.
●
Accumulation
Type
Average
●
Averaging 100
wfms
Содержание BERTWave MP2100B
Страница 24: ...xxiv...
Страница 96: ...Chapter 2 Before Use 2 42...
Страница 112: ...Chapter 3 Connecting with DUT 3 16...
Страница 250: ...Chapter 7 Measuring Waveform 7 64 Figure 7 12 1 Marker Display...
Страница 262: ...Chapter 7 Measuring Waveform 7 76...
Страница 276: ...Chapter 8 Operating Optical Interface 8 14...
Страница 304: ...Chapter 9 Performance Test 9 28...
Страница 313: ...10 4 Displaying Software Version 10 9 10 Maintenance 4 Touch Exit and then the software version display screen is closed...
Страница 320: ...Chapter 10 Maintenance 10 16...
Страница 374: ...Appendix E Performance Test Record Form E 10...
Страница 382: ...Index Index 6...