Chapter 6 Measuring Bit Error
6-20
CAUTION
●
When attenuators are connected to both the Data In and
Data In connectors, use attenuators with the same
attenuation. If the amount of attenuation is different, the
displayed threshold voltage and the actual threshold
voltage may be different.
●
If the input voltage to the attenuators is either more
than 5 V or less than –5 V, check that the power
consumed by the attenuators does not exceed the rated
power.
●
Check that the voltage of the signal attenuated by the
attenuators does not exceed the voltage displayed at
the Data In and Data In connectors.