Chapter 7 Measuring Waveform
7-22
Upon completion of the calibration, a dialog box appears and gives you the
calibration result.
Figure 7.5.1-1 Calibration Result Display
(When Calibration Is Successful)
Figure 7.5.1-2 Calibration Result Display (When Calibration Fails)
If the calibration fails, follow the instruction that applies to your situation,
and then retry from step 2.
●
Check that the coaxial cable is not connected to any of the Ch A In, Ch
B In and Trigger Clk In connectors on the front panel.
●
If it is required to keep the coaxial cables connected to the Ch A In, Ch
B In and Trigger Clk In connectors on the front panel, check that the
connectors do not receive any incoming signal.
When the Calibration of CHA and CHB Fails
When the Calibration of CHA Fails
When the Calibration of CHB Fails