6.1 Function List
The bit error rate can be measured at the Error Detector (ED).
The ED has the following functions.
Table 6.1-1 Items Set at ED
Item Description
Tracking
Sets test pattern setting at PPG.
Data Input Condition
Sets connector used for inputting signal.
Threshold
(Data Input Condition)
Sets threshold voltage for evaluating “0” and “1” at
Data
In and Data In
connectors.
The setting voltage range varies depending on the value of the External
Attenuator Factor.
External ATT
When inserting attenuator at connector, that attenuation is entered.
Test Pattern
Selects PRBS or Programmable Pattern.
POS/NEG
Sets the signal level of the Data connector responding to Pattern 1.
Programmable Pattern Selects the pattern file name to be used when Test Pattern is set to
Programmable Pattern. The file name selected is displayed.
Data Length
Displays pattern length in bit units.
Auto SYNC
Sets whether or not to execute resynchronization processing when pattern
sync lost
Threshold (Auto SYNC) When Auto SYNC is On, sets bit error rate judging pattern sync loss.
SYNC Control
When the pattern is Programmable Pattern, set the synchronization
method.
Frame Position
At Programmable Pattern, sets bit string header position used in sync
frame.
ED Result
Sets the method for displaying the bit error measurement results.
History Reset
Deletes history of warning display.
Gating Cycle
Sets single measurement, repeat measurement or continuous measurement.
Measurement Cycle
Sets the measurement time.
Current
Setting to [On] updates the results display while measurement is in
progress.
Setting to [Off] updates the results display after the measurement cycle has
elapsed.
Start/Stop
Stops/starts bit error rate measurement.
Содержание BERTWave MP2100B
Страница 24: ...xxiv...
Страница 96: ...Chapter 2 Before Use 2 42...
Страница 112: ...Chapter 3 Connecting with DUT 3 16...
Страница 250: ...Chapter 7 Measuring Waveform 7 64 Figure 7 12 1 Marker Display...
Страница 262: ...Chapter 7 Measuring Waveform 7 76...
Страница 276: ...Chapter 8 Operating Optical Interface 8 14...
Страница 304: ...Chapter 9 Performance Test 9 28...
Страница 313: ...10 4 Displaying Software Version 10 9 10 Maintenance 4 Touch Exit and then the software version display screen is closed...
Страница 320: ...Chapter 10 Maintenance 10 16...
Страница 374: ...Appendix E Performance Test Record Form E 10...
Страница 382: ...Index Index 6...