Chapter 3 Connecting with DUT
3-6
When the DUT is an optical transmitter:
Use the Option 051 and follow the connection procedures.
1. Connect
the
PPG2
Data
Out and Data Out to the DUT input
terminal using a coaxial cable.
2. Connect the DUT output connector to optical transceiver for an
optical input connector using an optical fiber.
3. Connect the Rx Data Out to ED2 Data In using a coaxial cable.
Do not connect anything to
Data
In
4. Connect the coaxial terminator provided with the BERTWave to
ED2
Data
In.
CAUTION
Check that the DUT optical output level does not exceed
the rated optical input level for the optical module inserted
with the Option 051.
There is a risk of damaging the optical module if the
optical power exceeding the rated optical input level is
applied.
PPG2
ED2
Optical Fiber
Coaxial Cable
Rx Data Out
Optical Input Connector
DUT
Figure 3.1-4 Connection when DUT is Optical Transmitter