Chapter 7 Measuring Waveform
7-20
7.4 Measurement Procedure
The following figure shows the basic measurement procedure.
Start
Input clock signal to
Trigger Clk In connector
Input measurement signal to
either Ch A or Ch B In*.
Set Bit Rate.
Calibration
required?
Perform calibration.
Set EYE/pulse mode.
Set Clock Frequency/Clock
Divide ratio.
Pulse mode?
Set pattern length.
Sets data acquired method.
Set Ch A or Ch B
display to On.
Start sampling.
Measure waveform.
End
Stop sampling.
Adjust scale.
Using clock
recovery unit
(CRU)?
Yes
No
Yes
No
Yes
No
Connect Trigger
Clk In and CRU
Out connectors
Save measurement
Figure 7.4-1 Basic Measurement Procedures