7.11 Measuring Waveform
7-49
7
Measurin
g W
aveform
To set measurement method of rise/fall time
When measuring the rise time and fall time levels, select from amplitude
10/90% level or 20/80% level.
●
Touch the Rise/Fall Time button to display [10/90%] or [20/80%].
The bandwidth of the sampling oscilloscope is corrected with the
measurement value of the Rise/Fall time, and the corrected bandwidth is
displayed.
1. Touch [Rise/Fall Time Correction] to set the button display to [On].
2. Touch the Correction Factor text box.
3. Input the correction value in ps.
The calculation formula is as follows:
2
2
Tc
Tm
Td
Td: display value (ps), Tm: measurement value (ps),
Tc: correction value (ps)
If the measurement value is larger than the correction value, the result
displays N/A.
"Corrected" is displayed in red at the Rise/Fall time when Rise/Fall Time
Correction is set to [On].
Figure 7.11.1-5 Display Example of Measurement Result