Chapter 9 Performance Test
9-6
19. Connect the PPG1 Data Out connector to the input connector
of the sampling oscilloscope.
20. Repeat steps 11 to 14.
PPG2 can be tested in the same way with the Option 012. The connection
diagram is as follows.
Sampling Oscilloscope
Sync Out
Trigger
Input
PPG2 Data Out
Input
Coaxial Terminator
Figure 9.1.3-3 PPG2 Data Out Waveform Test Connection Diagram
Sampling Oscilloscope
Sync Out
Trigger
Input
PPG2 Data Out
Input
Coaxial Terminator
Figure 9.1.3-4 PPG2
Data
Out Waveform Test Connection Diagram
ED3 and ED4 can be tested in the same way with the Option 014. The
connection diagram for measuring the ED3 input level is as follows.