6.3 Setting Conditions for Error Detection
6-17
6
Measuring Bit Error
The following table shows the sync processing start and stop levels for
each Sync Threshold selection.
Table 6.3-1 Bit Error Rate when Threshold setting is INT
Test Pattern
Sync Control
Pattern Length
Bit Error Rate
Sync
Processing
Start Level
Sync
Processing
Stop Level
PRBS – 2
n
–1 (n=7,9,15,23,31) 2.5 × 10
2
1.56 × 10
2
Programmable
Pattern
OFF
2 to 16
2.5 × 10
2
1.56 × 10
2
17 to 160
5.0 × 10
3
1.56 × 10
3
161 to 1,600
5.0 × 10
4
1.56 × 10
4
1,601 to 16,000
5.0 × 10
5
1.56 × 10
5
16,001 to 80,000
1.0 × 10
5
6.25 × 10
6
80,001 to 160,000
5.0 × 10
6
3.13 × 10
6
160,001 to 320,000
2.5 × 10
6
1.56 × 10
6
320,001 to 524,288
1.9 × 10
6
7.81 × 10
7
524,289 to 1,048,576
9.54 × 10
7
3.91 × 10
7
ON
128 to 5,120
1.95 × 10
4
1.56 × 10
4
5,121 to 10,240
9.77 × 10
5
7.81 × 10
4
10,241 to 51,200
1.95 × 10
5
1.56 × 10
5
51,201 to 102,400
9.77 × 10
6
7.81 × 10
5
102,401 to 204,800
4.88 × 10
6
3.91 × 10
6
204,801 to 307,200
3.26 × 10
6
2.60 × 10
6
307,201 to 409,600
2.44 × 10
6
1.95 × 10
6
409,601 to 524,288
1.91 × 10
6
1.53 × 10
6
524,289 to 1,048,576
9.54 × 10
7
7.63 × 10
7
Содержание BERTWave MP2100B
Страница 24: ...xxiv...
Страница 96: ...Chapter 2 Before Use 2 42...
Страница 112: ...Chapter 3 Connecting with DUT 3 16...
Страница 250: ...Chapter 7 Measuring Waveform 7 64 Figure 7 12 1 Marker Display...
Страница 262: ...Chapter 7 Measuring Waveform 7 76...
Страница 276: ...Chapter 8 Operating Optical Interface 8 14...
Страница 304: ...Chapter 9 Performance Test 9 28...
Страница 313: ...10 4 Displaying Software Version 10 9 10 Maintenance 4 Touch Exit and then the software version display screen is closed...
Страница 320: ...Chapter 10 Maintenance 10 16...
Страница 374: ...Appendix E Performance Test Record Form E 10...
Страница 382: ...Index Index 6...