Chapter 6 Measuring Bit Error
6-24
6.5 Measurement Result
The following measurement result is displayed in the ED Result.
Start Time:
Time when bit error measurement started
Elapsed Time: Time elapsed from start of bit error rate measurement
When Gating Cycle is Single or Repeat, when time exceeds
the Time setting at Gating, the time is reset back to 0.
Remaining Time: Bit error measurement elapsed time subtracted from
time set at Gating Time setting.
ER: The bit error rate is displayed from 0.0001E–18 to 1.0000E–0.
Mantissa is 0.0000 when the bit error does not occur.
At this time, the numeric part varies depending on the clock number.
Example: 0.0000E–3
Clock count 1000 or more, 9999 or less
0.0000E–4
Clock count 10000 or more, 99999 or less
EC: The occurred bit errors are displayed from 0 to 9999999 or from
1.0000E07 to 9.9999E17.
CC: The received bit count is displayed from 0 to 9999999 or from
1.0000E07 to 9.9999E17.
FREQ (kHz): Clock frequency calculated by the received bit count.
It is the same as the transmission speed (kbit/s) of the received data.
Alarm Indicator
Error: Displays
red
when
detecting bit error.
SYNC Loss: Displays red when pattern not synchronized.
When the alarm indicator goes red once, the History Indicator goes yellow
when the cause of the alarm is removed, indicating that an alarm
occurred.
When the [History Reset] button is pressed when the History Indicator is
yellow, the yellow display disappears.
Содержание BERTWave MP2100B
Страница 24: ...xxiv...
Страница 96: ...Chapter 2 Before Use 2 42...
Страница 112: ...Chapter 3 Connecting with DUT 3 16...
Страница 250: ...Chapter 7 Measuring Waveform 7 64 Figure 7 12 1 Marker Display...
Страница 262: ...Chapter 7 Measuring Waveform 7 76...
Страница 276: ...Chapter 8 Operating Optical Interface 8 14...
Страница 304: ...Chapter 9 Performance Test 9 28...
Страница 313: ...10 4 Displaying Software Version 10 9 10 Maintenance 4 Touch Exit and then the software version display screen is closed...
Страница 320: ...Chapter 10 Maintenance 10 16...
Страница 374: ...Appendix E Performance Test Record Form E 10...
Страница 382: ...Index Index 6...