A.1 MP2100B BERTWave
A-17
A
pp
en
dix
A
ppe
ndi
x A
In this example, the difference between 1 level voltage (200 mV) and
offset voltage (50 mV) is 150 mV. At a scale of 50 mV/div for a voltage
difference of 150 mV, the amplitude accuracy is 8 mV. At 1 level accuracy,
the difference is calculated as follows:
200 × 2% + 8 = 12
±12 mV
The difference between 0 level voltage (–200 mV) and offset voltage (50
mV) is 250 mV.
At the 50 mV/div scale for a voltage difference of 250 mV, the amplitude
accuracy is 11 mV. At 0 level accuracy, the difference is calculated as
follows:
200 × 2% + 11 = 15
±15 mV