5-33
Troubleshooting
Routine Maintenance
129367-410
STATUS
PAPER
FEED
SILENCE
ALARM
STATUS..
ALARM! LOAD
286
DOOR A SEAL SWITCH MALFUNCTION
CHAMBER:
104.3 F
5.4 psig
STERILIZER WILL:
•
AUTOMATICALLY ABORT CYCLE
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. CALL SERVICE
Door Seal A Switch
Malfunction
Occurs if switched
neutral relay associated
with door seal A, closed
switch fails.
STATUS
PAPER
FEED
SILENCE
ALARM
STATUS..
ALARM! LOAD
280
BOARD OVER TEMP FAILURE
CHAMBER:
300.0 F
0.0 psig
STERILIZER WILL:
•
AUTOMATICALLY ABORT CYCLE
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. CALL SERVICE
Board Over Temp Failure
Occurs if temperature
of the main control
board exceeds
acceptable conditions.
Alarm
Description
Screen with Operator Instructions
STATUS..
SERVICE INFORMATION:
287
DOOR A SEAL SWITCH MALFUNCTION
→
SEAL SWITCH CLOSED OUT OF CYCLE
CAUSES AND CORRECTION:
1.
SEAL SWITCH MALFUNCTION
→
CHECK LPS1 (PS2) FOR FAILED CLOSED
2.
SHORTED CABLE
→
CHECK CABLE AT P19 FOR SHORT CIRCUIT
CLEAR
PAPER
EXIT
ALARM
FEED
Screen with Operator Instructions
Screen with Operator Instructions