5-26
129367-410
Routine Maintenance
Troubleshooting
Alarm
Description
Screen with Service Instructions
Screen with Operator Instructions
STATUS..
SERVICE INFO: CHAMBER
218
PRESSURE TRANSDUCER FAILURE
→
TRANSDUCER, CP, OUTPUT VOLTAGE IS
OUTSIDE NORMAL RANGE
CAUSES AND CORRECTION:
1.
LOOSE CONNECTION IN TRANSDUCER
WIRING
→
REPAIR
2.
TRANSDUCER FAILED
→
REPAIR
→
RECALIBRATE
3.
CONTROL OUT OF CALIBRATION
→
RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
4.
MAIN CONTROL FAILURE
→
CHECK CONTROL BOARD POWER STATUS
LEDs
→
REPLACE 146656-185 CONTROL BOARD
→
RECALIBRATE
PAPER
SERVICE
EXIT
FEED
MODE
Chamber Pressure
Transducer Failure,
Continued
Occurs if chamber
temperature reading is
outside the normal
range.
Chamber Temperature
Probe Failure
STATUS
PAPER
FEED
SILENCE
SERVICE
ALARM
HELP
STATUS..
ALARM! CHAMBER
211
TEMPERATURE PROBE FAILURE
CHAMBER:
000.0 C
00.0 psig
STERILIZER WILL:
•
AUTOMATICALLY ABORT CYCLE
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. CALL SERVICE
For Life Sciences
Only
STATUS..
SERVICE INFO: CHAMBER
218
PRESSURE TRANSDUCER FAILURE
→
TRANSDUCER, CP, OUTPUT VOLTAGE IS
OUTSIDE NORMAL RANGE
CAUSES AND CORRECTION:
1.
LOOSE CONNECTION IN TRANSDUCER
WIRING
→
REPAIR
2.
TRANSDUCER FAILED
→
REPAIR
→
RECALIBRATE
3.
CONTROL OUT OF CALIBRATION
→
RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
4.
MAIN CONTROL FAILURE
→
CHECK CONTROL BOARD POWER STATUS
LEDs
→
REPLACE CONTROL BOARD*
→
RECALIBRATE
PAPER
EXIT
FEED
For Healthcare
Only
*Depending upon the configura-
tion of the unit, screen text
may reference 146656-185
control board, or 146659-065
control board.