5-6
129367-410
Routine Maintenance
Troubleshooting
Alarm
Description
Screen with Service Instructions
Too Long In Evacuation,
Continued
ABORT
STATUS
PAPER
MORE
FEED
HELP
STATUS..
SERVICE INFORMATION:
203
TOO LONG IN EVACUATION
→
CHAMBER DID NOT REACH REQUIRED
VACUUM LEVEL WITHIN ALLOTTED TIME
CAUSES AND CORRECTION:
1.
WATER PRESSURE LESS THAN 30 PSIG
→
CHECK WATER SUPPLY PIPING
2.
CHAMBER DRAIN STRAINER PLUGGED
→
CLEAN
3.
CHECK VALVE MALFUNCTION
→
REPAIR
4.
SOLENOID VALVE MALFUNCTION
→
REPAIR
5.
DOOR SEAL NOT ACTIVATED
→
CHECK SEAL
→
CHECK SEAL STEAM AND EXHAUST
For Healthcare
Only
→
CHAMBER DID NOT REACH REQUIRED
VACUUM LEVEL WITHIN ALLOTTED TIME
CAUSES AND CORRECTION:
6.
LEAK IN PLUMBING
→
REPAIR
→
RUN A LEAK TEST
7.
CONTROL OUT OF CALIBRATION
→
RECALIBRATE CONTROL (CONTACT
QUALIFIED SERVICE PERSON)
STATUS..
SERVICE INFORMATION:
204
TOO LONG IN EVACUATION
PAPER
SERVICE
EXIT
FEED
MODE
For Life Sciences
Only
→
CHAMBER DID NOT REACH REQUIRED
VACUUM LEVEL WITHIN ALLOTTED TIME
CAUSES AND CORRECTION:
6.
LEAK IN PLUMBING
→
REPAIR
→
RUN A LEAK TEST
7.
CONTROL OUT OF CALIBRATION
→
RECALIBRATE CONTROL (CONTACT
QUALIFIED SERVICE PERSON)
STATUS..
SERVICE INFORMATION:
204
TOO LONG IN EVACUATION
PAPER
EXIT
FEED
For Healthcare
Only