5-32
129367-410
Routine Maintenance
Troubleshooting
Alarm
Description
Screen with Operator Instructions
STATUS
PAPER
FEED
SILENCE
ALARM
STATUS..
ALARM! LOAD
277
ROM FAILURE
CHAMBER:
300.0 F
0.0 psig
STERILIZER WILL:
•
AUTOMATICALLY ABORT CYCLE
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. CALL SERVICE
ROM Failure
Occurs if read-only
memory on the main
control board fails.
STATUS
PAPER
FEED
SILENCE
ALARM
STATUS..
ALARM! LOAD
278
RAM FAILURE
CHAMBER:
300.0 F
0.0 psig
STERILIZER WILL:
•
AUTOMATICALLY ABORT CYCLE
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. CALL SERVICE
RAM Failure
Occurs if random-
access memory on the
main control board fails.
STATUS
PAPER
FEED
SILENCE
ALARM
STATUS..
ALARM! LOAD
279
ADC FAILURE
CHAMBER:
300.0 F
0.0 psig
STERILIZER WILL:
•
AUTOMATICALLY ABORT CYCLE
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. CALL SERVICE
ADC Failure
Occurs if analog-to-
digital board on the
main control board fails.
Screen with Operator Instructions
Screen with Operator Instructions