5-29
Troubleshooting
Routine Maintenance
129367-410
Alarm
Description
Screen with Operator Instructions
Occurs if door seal
switch contact is made
but door switch is still
open.
Door Switch Failure
STATUS
PAPER
FEED
SILENCE
SERVICE
ALARM
HELP
STATUS..
ALARM!
237
DOOR SWITCH FAILURE (LS1)
CHAMBER:
000.0 C
00.0 psig
STERILIZER WILL:
•
AUTOMATICALLY ABORT CYCLE
•
EXHAUST CHAMBER TO ATMOSPHERIC
PRESSURE
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. CALL SERVICE
Screen with Service Instructions
STATUS..
SERVICE INFORMATION:
238
DOOR SWITCH FAILURE
→
DOOR SWITCH OPEN WHILE SEAL SWITCH
CLOSED
CAUSES AND CORRECTION:
1.
DOOR SWITCH MALFUNCTION
→
CHECK LS1(LS2) CONNECTIONS
→
READJUST LS1(LS2)
→
REPAIR LS1(LS2)
2.
SEAL PRSSURE SWITCH MALFUNCTION
→
READJUST PS1(PS2)
→
REPAIR PS1(PS2)
PAPER
SERVICE
EXIT
FEED
MODE
For Life Sciences
Only
For Healthcare
Only
STATUS..
SERVICE INFORMATION:
238
DOOR SWITCH FAILURE
→
DOOR SWITCH OPEN WHILE SEAL SWITCH
CLOSED
CAUSES AND CORRECTION:
1.
DOOR SWITCH MALFUNCTION
→
CHECK LS1CONNECTIONS
→
READJUST LS1
→
REPAIR LS1
CLEAR
PAPER
EXIT
ALARM
FEED