5-23
Troubleshooting
Routine Maintenance
129367-410
Alarm
Description
Screen with Operator Instructions
Screen with Service Instructions
Occurs if door seal
does not reach 10 psig
within allotted time.
Too Long To Seal Door
STATUS
PAPER
FEED
SILENCE
SERVICE
ALARM
HELP
STATUS..
ALARM!
227
TOO LONG TO SEAL DOOR
CHAMBER:
000.0 C
00.0 psig
STERILIZER WILL:
•
AUTOMATICALLY ABORT CYCLE
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. CHECK STEAM SUPPLY VALVE
→
IF CLOSED, OPEN VALVE AND
RESELECT CYCLE
3. IF ALARM RECURS, CALL SERVICE
STATUS..
SERVICE INFORMATION:
228
TOO LONG TO SEAL DOOR
→
DOOR SEAL DID NOT REACH 10 PSIG
WITHIN ALLOTTED TIME
CAUSES AND CORRECTION:
1.
SEAL PRESSURE SWITCH MALFUNCTION
→
READJUST PS1(PS2)
→
REPAIR PS1(PS2)
2.
SEAL NOT ACTIVATING
→
CHECK SEAL STEAM
→
CHECK SEAL EXHAUST
3.
SOLENOID VALVE MALFUNCTION
→
REPAIR S35(S36)
PAPER
CLEAR
EXIT
FEED
ALARM
For Life Sciences
Only
For Life Sciences
Only
STATUS
PAPER
FEED
SILENCE
ABORT
SERVICE
ALARM
HELP
STATUS..
ALARM!
227
TOO LONG TO SEAL DOOR
CHAMBER:
000.0 C
00.0 psig
STERILIZER WILL:
•
AUTOMATICALLY ABORT CYCLE
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. CHECK STEAM SUPPLY VALVE
→
IF CLOSED, OPEN VALVE AND
RESELECT CYCLE
3. IF ALARM RECURS, CALL SERVICE
For Healthcare
Only