5-2
129367-410
Routine Maintenance
Troubleshooting
SILENCE
ABORT
SERVICE
ALARM
HELP
For Healthcare
Only
CHAMBER:
000.0 C
00.0 psig
STERILIZER WILL:
•
AUTOMATICALLY TRY TO COMPLETE CYCLE
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. CHECK STEAM SUPPLY VALVE
→
IF CLOSED, OPEN VALVE
3. IF ALARM RECURS, CALL SERVICE
STATUS..
ALARM!
205
TOO LONG IN CHARGE
Figure 5-1. Typical Alarm Screen
Current Chamber Status
Alarm Name
Operator Instructions
Screen
Reference
Number
Sterilizer Activity
STATUS
PAPER
FEED
5.1.2 Typical Alarm
Printout
* ALARM
PRESSURE IN CHAMBER
F 10:07:23A
61.7C
34.OP
Full Print Format Shown
Figure 5-2. Typical Alarm
Printout
Touch screen buttons, located along bottom of alarm screen, are used to
perform the following functions.
•
Pressing
SILENCE ALARM
turns off the alarm buzzer.
•
Pressing
STATUS PRINT
generates a printout of the current temperature
and pressure in the sterilizer chamber at the time the button was pressed.
•
Pressing
PAPER FEED
advances the printer paper up by one line.
•
Pressing
SERVICE HELP
advances display to the corresponding service
information screen. This screen provides the qualified service technician
with possible causes and advanced corrective actions for that alarm
condition.
IMPORTANT: In the event of an alarm condition, the operator should always
follow the instructions indicated on the alarm screen.
When an alarm occurs the printer automatically generates a printout, typically
listing alarm name, time alarm occurred, current chamber status and any
associated sensor temperature. See Figure 5-2.
The following alarm screens will appear only during cycle operation.
5.2 In-Cycle Alarms
Alarm
Description
Screen with Operator Instructions
Occurs if chamber does
not reach the set tem-
perature within the
allotted time.
Too Long In Charge
STATUS
PAPER
FEED
SILENCE
SERVICE
ALARM
HELP
CHAMBER:
000.0 C
00.0 psig
STERILIZER WILL:
•
AUTOMATICALLY TRY TO COMPLETE CYCLE
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. CHECK STEAM SUPPLY VALVE
→
IF CLOSED, OPEN VALVE
3. IF ALARM RECURS, CALL SERVICE
STATUS..
ALARM!
205
TOO LONG IN CHARGE
For Life Sciences
Only