5-9
Troubleshooting
Routine Maintenance
129367-410
Alarm
Description
Screen with Service Instructions
Under Sterilize
Temperature,
Continued
STATUS..
SERVICE INFORMATION:
224
UNDER STERILIZE TEMPERATURE
→
CHAMBER TEMPERATURE DROPPED BELOW
STERILIZE TEMPERATURE BY UNDERTEMP
TEMPERATURE VALUE
CAUSES AND CORRECTION:
1.
STEAM PRESSURE LESS THAN 50 PSIG
→
CHECK STEAM SUPPLY PIPING
2.
CHAMBER STEAM TRAP MALFUNCTION
→
REPAIR
3.
SOLENOID VALVE MALFUNCTION
→
REPAIR S09
→
REPAIR S02
4.
CONTROL OUT OF CALIBRATION
→
RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
ABORT
PAPER
EXIT
FEED
NOTE: If under-
temperature value is
set to "RESUME", the
sterilizer will auto-
matically resume
sterilize time after set
temperature is
reached.
For Healthcare
Only
For Life Sciences
Only
Occurs if chamber
temperature exceeds
the maximum sterilize
temperature (control
temperature plus over
temperature value).
SILENCE
SERVICE
ALARM
HELP
CHAMBER:
000.0 C
00.0 psig
STERILIZER WILL:
•
AUTOMATICALLY CONTINUE CYCLE
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. IF ALARM RECURS, ABORT CYCLE
AND CALL SERVICE
STATUS..
ALARM!
235
OVER STERILIZE TEMPERATURE
ABORT
Over Sterilize
Temperature
CHAMBER:
000.0 C
00.0 psig
STERILIZER WILL:
•
AUTOMATICALLY CONTINUE CYCLE
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. IF ALARM RECURS, ABORT CYCLE
AND CALL SERVICE
STATUS..
ALARM!
235
OVER STERILIZE TEMPERATURE
For Healthcare
Only
SILENCE
ABORT
SERVICE
ALARM
HELP
STATUS
PAPER
FEED
Screen with Operator Instructions