5-21
Troubleshooting
Routine Maintenance
129367-410
The following alarm screens will appear any time the sterilizer is energized.
The sensors are continually monitored whenever the sterilizer is in or out of
cycle.
5.4 Sensor Alarms
Alarm
Description
Screen with Operator Instructions
Water In Chamber
Occurs if excess water
is sensed in the
chamber.
STATUS
PAPER
FEED
SILENCE
SERVICE
ALARM
HELP
CHAMBER:
000.0 C
00.0 psig
WARNING - BURN HAZARD!
CHAMBER MAY BE FILLED WITH STEAM
CONDENSATE
OPERATOR INSTRUCTIONS:
1. DO NOT OPEN DOOR
2. SILENCE ALARM
3. CALL SERVICE IMMEDIATELY
STATUS..
ALARM!
209
WATER IN CHAMBER
Screen with Service Instructions
STATUS..
SERVICE INFORMATION:
210
WATER IN CHAMBER
→
EXCESS WATER SENSED IN CHAMBER,
WARNING! BURN HAZARD
CAUSES AND CORRECTION:
1.
JACKET STRAINER PLUGGED
→
CLEAN
2.
JACKET TRAP FAILED CLOSED
→
REPAIR
3.
CHAMBER TRAP FAILED CLOSED
→
REPAIR
4.
WATER ENTERED CHAMBER THROUGH STEAM
PIPING
→
CHECK BOILER OR STEAM GENERATOR
→
REPAIR PIPING
5.
WATER FLOAT SENSOR MALFUNCTION
→
REPAIR
PAPER
SERVICE
EXIT
FEED
MODE
For Life Sciences
Only
STATUS..
SERVICE INFORMATION:
210
WATER IN CHAMBER
→
EXCESS WATER SENSED IN CHAMBER,
WARNING! BURN HAZARD
CAUSES AND CORRECTION:
1.
JACKET STRAINER PLUGGED
→
CLEAN
2.
JACKET TRAP FAILED CLOSED
→
REPAIR
3.
CHAMBER TRAP FAILED CLOSED
→
REPAIR
4.
WATER ENTERED CHAMBER THROUGH STEAM
PIPING
→
CHECK BOILER OR STEAM GENERATOR
→
REPAIR PIPING
5.
WATER FLOAT SENSOR MALFUNCTION
→
REPAIR
PAPER
EXIT
FEED
For Healthcare
Only