5-10
129367-410
Routine Maintenance
Troubleshooting
Alarm
Description
Screen with Service Instructions
Screen with Operator Instructions
Occurs if cycle remains
in the sterilize phase
longer than the allotted
time. (Lab/Sci program
option only.)
Too Long In Sterilize
STATUS
PAPER
FEED
SILENCE
SERVICE
ALARM
HELP
STATUS..
ALARM!
233
TOO LONG IN STERILIZE
CHAMBER:
000.0 C
00.0 psig
STERILIZER WILL:
•
AUTOMATICALLY ABORT CYCLE
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. CALL SERVICE
Over Sterilize
Temperature
Continued
STATUS..
SERVICE INFORMATION:
236
OVER STERILIZE TEMPERATURE
→
STERILIZE TEMPERATURE IS ABOVE
SETPOINT BY MORE THAN PRESCRIBED
AMOUNT
CAUSES AND CORRECTION:
1.
STEAM PRESSURE MORE THAN 50 PSIG
→
CHECK STEAM SUPPLY PIPING
2.
CHAMBER STEAM TRAP MALFUNCTION
→
REPAIR
3.
SOLENOID VALVE MALFUNCTION
→
REPAIR S09
→
REPAIR S02
4.
CONTROL OUT OF CALIBRATION
→
RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
SERVICE
STATUS
PAPER
EXIT
MODE
FEED
For Life Sciences
Only
STATUS..
SERVICE INFORMATION:
236
OVER STERILIZE TEMPERATURE
→
STERILIZE TEMPERATURE IS ABOVE
SETPOINT BY MORE THAN PRESCRIBED
AMOUNT
CAUSES AND CORRECTION:
1.
STEAM PRESSURE MORE THAN 50 PSIG
→
CHECK STEAM SUPPLY PIPING
2.
CHAMBER STEAM TRAP MALFUNCTION
→
REPAIR
3.
SOLENOID VALVE MALFUNCTION
→
REPAIR S09
→
REPAIR S02
4.
CONTROL OUT OF CALIBRATION
→
RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
For Healthcare
Only
PAPER
EXIT
FEED